Safety Tester
ASA Housing LED Flood light 50W
Information d Electrical Safety Comprehensive tester

AN9652S(F)/AN9652F(F),AC Hipot,Insulation Resistance,Ground Bond resistance,Leakage Current,Running Power,Low-Voltage Start.

  • Details
  • Sales area
  • Enquiry


 

  THE SAFETY TESTER IS THE GENERIC TERM TO COVER THE 3 OR 8 TESTS OF THE ELECTRICAL SAFETY STANDARDS, FROM THE INTERNATIONAL STANDARDS

  ground bond test (also called bonding test, ground continuity test )

  - insulation test (also called high resistance test, megger test)

  - hipot test to make a high voltage test (also called dielectric strength test, hipot test, hi-pot test, breakdown test or flash test)

  - leakage current measurement (also called line leakage test)

  The final purpose of those tests is to validate that the product under test is safe for the end user, and complies with the quality requirements from the Low Voltage Directive (LVD), applicable to the corresponding standards (IEC335, IEC950, IEC601, IEC1010, IEC598, IEC990, IEC204...)

  APPLICATION

  MANUFACTURER, LABORATORY, HIGH SPEED PRODUCTION, MAINTENANCE, IN ORDER TO COMPLY WITH THE LOW VOLTAGE DIRECTIVE

  

  • IEC 335 household manufacturer (washing machines, coffee machines, fridges, heaters, boilers, …)

  • IEC 204 industrial equipment (compressor, electrical tools, …)

  • IEC 598 lighting industry (street lights, professional lighting, medical lighting, bulbs, …)

  • IEC 601 medical industry (scanners, surgery equipments, X-ray, electrical beds, …)

  • IEC 950 information technology industry (modems, satellite receivers, servers, printers,cash dispensers, …)

  • IEC 1010 measurement, control and laboratories industry (electrical meters, test & measurement equipment, …)

  SPECIFICATION OVERVIEW

  FROM LABORATORY TO PRODUCTION TESTING, OUR RANGE COVERS THE FOLLOWING SPECIFICATIONS

  • ground bond test up to 64A AC with resistance measurement up to 600mΩ

  • insulation test with DC Voltage from 1V up to 1000 VDC (1V steps),500VA AC Hipot capacity(AC 5KV/100mA),short-circuit output current exceeds 200mA.Built-in optional leakage current test network MD card,meeting requirements of different industrial standards Maximum test current of Ground Bond resistance is up to 64A

  • leakage current measurement in single or three phases, according to all current international standards

  • AC Hipot,Insulation Resistance,Ground Bond resistance,Leakage Current,Running Power,Low-Voltage Start.

  • power measurement and cos φ measurement

  • Satisfy lattest requirements:Follow the production line developing trends of comprehensive,fast,automation and information.Satisfy quality management,5s management and refid management requirements.

  • GPIB, RS232, Ethernet, 0-10V, PLC available

  • ESRS software for quality purposes, traceability and professional reporting

Technical specification:

Model AN9652S/AN9652F
Rated output 200VA (5000V/40mA), short circuit current is greater than 200mA;
[Optional: 5000V/100mA]
Output voltage Range/accuracy 100~5000V, ± (2%× setting + 5 words)
Output frequency Range/accuracy 50Hz or 60Hz, ±0.1%× setting
Alarm current Upper limit/accuracy 0.10~ or 40.00mA, ± (2%× setting + 5 words)
Lower limit /accuracy 0.000~9.999mA, ± (2%× setting + 5 words)
Time Range 0.5~999.9s (0: infinite time)
Ramp time 0.1~999.9s
Descent time 0.1~999.9s
Accuracy ± (0.2%× setting + 1 word)
Measure meter Voltage range/accuracy 0.01~5.00kV, ± (1.5%× reading + 1 word)
Frequency range/accuracy 50Hz or 60Hz, ±0.1%× reading
Current range/accuracy 0.10~40.00mA, ± (2%× reading + 2 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading + 1 word)
Output voltage wave, distortion, adjusting rate Sine wave, less than 2% (pure resistive load), less than (2% × setting + 5V) (from zero to full load)
Start voltage (0%~50%)× output voltage setting
Current compensation 0.000~40.00mA, Auto/manual
Arc detection Level 1~9, 9: highest level; 0: off
DC withstand voltage test
Output voltage Range/accuracy 100~6000Vdc, ± (2%× setting + 5 words)
Alarm current Upper limit/accuracy 0~10000μA, ± (2%× setting + 5 words)
Lower limit /accuracy 0.0~999.9μA, ± (2%× setting + 5 words)
Time Range 0.5~999.9s (0: infinite time)
Ramp time 0.1~999.9s
Descent time 0.3~999.9s (0, off)
Accuracy ± (0.2%× setting + 1 word)
Measure meter Voltage range/accuracy 0.01~6.00Vdc, ± (1.5%× reading + 1 word)
Current range/accuracy 0.0~10000uA, ± (2%× reading + 5 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading + 1 word)
Rated output 60VA (6000Vdc/10mA)
Output voltage fluctuation, regulation Less than 5% (purely resistive load, 6000V/10mA), less than (2% × setting + 5V) (from zero to full load)
Compensation current 0 ~200.0μA, Auto/manual
Start voltage (0%~50%)×output voltage setting
Charging current lower limit (0~3.500) μA, Auto/manual
Discharge time ≤200ms
Arc detection Level 1~9, 9: highest level; 0: off
Insulation resistance test
Output voltage Range/accuracy 100~2500Vdc, ± (2%× setting +5V)
Alarm resistance Upper/lower limit 1~9999MΩ
Accuracy 100-499V, 1~1000MΩ, ± (8%× setting + 2 words)
1000~2000MΩ, ± (12%× setting + 2 words)
500-2500V, 1~199MΩ, ± (2%× setting + 2 words),
200~999MΩ, ± (5%× setting + 2 words)
1000~9999MΩ, ± (15%× setting + 2 words)
Time Ramp time 0.1~999.9s
Delay decision time 0, 0.5~999.9s (0: infinite time)
Range 0, 0.5~999.9s (0: infinite time)
Descent time 0.1~999.9s
Accuracy ± (0.2%× setting + 1 word)
Measure meter Voltage range/accuracy 100~2500Vdc, ± (2%× reading + 5 words)
Resistance range/accuracy 1~9999MΩ;
100-499V, 1~1000MΩ, ± (8%× reading + 2 words)
1000~2000MΩ, ± (12%× reading + 2 words)
500-2500V, 1~199MΩ, ± (2%× reading + 2 words),
200~999MΩ, ± (5%× reading + 2 words)
1000~9999MΩ, ± (15%× reading + 2 words)
Time range/accuracy 0.1~999.9s, ± (0.2%× reading + 1 word)
Charging current lower limit 0~3.500μA, Auto/manual
Grounding resistance test
Rated output Maximum output current: 32A; maximum test resistance: 600mΩ; maximum open circuit voltage: 12V;
[Optional: 64A ground test current]
Output current Range/accuracy 2.0~32.0A, ± (2%× setting + 5 words)
Output voltage Range/accuracy 3.0~10.0V, ± (2%× setting + 5 words)
Output frequency Range/accuracy Sine wave, 50Hz or 60Hz, ±0.1%× setting
Alarm limit Resistance upper/lower limit 2.0~10.6A: 0.1~600mΩ;
10.7~32.0A: 0.1~R mΩ, R= (6400/current setting) mΩ. As the current range is switched, the tester will automatically calculate the maximum range.
Error ± (2%× setting + 5 words)
Compensation resistance Range 0~100 mΩ, automatic measurement , compensation On/Off
Test time Range/accuracy 0.5~999.9s (0: infinite time); ± (0.2%× setting + 1 word)





Measure meter
Current range/accuracy 2.0~32.0A ( or 64A), ± (2%× reading + 2 words)
Voltage range/accuracy 3.0~10.0V , ± (2%× reading + 5 words)
Resistance range/accuracy 10.0~99.9~100~600mΩ; ± (2%× reading + 5 words)
Current Precision

Resistance
Time range/accuracy 0.1~999.9s; ± (0.2%× setting + 1 word)
Leakage Current Test
Type Single-phase load (300V/20A), dynamic leakage current test under the operating temperature and static leakage current test under the non-operating temperature
Leak test body network (MD card) Standard network, GB/T 12113 Fig. 4; [Optional: other leak test network]
Alarm leakage current Upper limit 0.050~9.999mA
Lower limit 0.000~5.000mA
Leakage current compensation Range 0.000~1.000mA, automatic measurement , compensation On/Off
Test time Range/accuracy 0.5~999.9s (0: infinite time), ± (0.2%× setting + 1 word)
Leakage current measurement Range 0.050~9.999mA
Accuracy DC~10kHz: ± (3%× reading +0.010mA)
10k~1MHz: ± (5%× reading +0.050mA)
Output voltage measurement Range/accuracy 60~280V (phase voltage), ± (0.5%× reading + 2 words)
Built-in programmable inverter power supply
Maximum output 5000VA
Output voltage distortion 1%@ 47-63Hz
Output voltage stability 1%
Programmable output voltage Range 0-300V
Accuracy 0.5%RD+0.5%F.S.
Programmable output frequency Range 47-63Hz
Accuracy 0.1%
Maximum output current 0-150V 27A
0-300V 14A
General Specification
Verification & Calibration 23℃±5℃, (45~75) %RH , less dust
Ambient 0 ºC ~40ºC, , (5~90) %RH , less dust
Storage -10℃~55℃, (5~95) %RH , less dust
Power supply 220V±10%, 50Hz±5%
Power consumption Maximum 1000VA (excluding operating power of objects under test)
Weight (without packaging) Approx. 80kg
Dimensions (without packaging, mm) 483 (W)x 600 (D)x 1355 (H)

 


 India  Malaysia  Indonesia
 Brazil  Venezuela  Thailand
 Poland  Hungary  Czech Republic
 Bangladesh  Sri Lanka  Vietnam
 Turkey  Egypt  Pakistan
 Korea  Australia  New-Zealand
 Russia  Saudi Arabia  Dubai
 United States  Canada  Mexico
 Nicaragua  Haiti Columbia
 Ecuador  Peru  Uruguay
Argentina  Chile  Italy
Austria  Czech Rep  Belgium
Netherlands United Kingdom Ireland
France Spain Germany
Portugal Finland  

Product:
Country:
Company:
Name:
Phone:
Email:
Message:
 
Products / Products