Stand Electrical Safety Comprehensive tester

AN9651S(F)/AN9651F(F)/AN9651B(F),AC Hipot,Insulation Resistance,Ground Bond resistance,Leakage Current,Running Power,Low-Voltage Start.

  • Details
  • Sales area
  • Enquiry

  THE SAFETY TESTER IS THE GENERIC TERM TO COVER THE 3 OR 8 TESTS OF THE ELECTRICAL SAFETY STANDARDS, FROM THE INTERNATIONAL STANDARDS

  ground bond test (also called bonding test, ground continuity test )

  - insulation test (also called high resistance test, megger test)

  - hipot test to make a high voltage test (also called dielectric strength test, hipot test, hi-pot test, breakdown test or flash test)

  - leakage current measurement (also called line leakage test)

  The final purpose of those tests is to validate that the product under test is safe for the end user, and complies with the quality requirements from the Low Voltage Directive (LVD), applicable to the corresponding standards (IEC335, IEC950, IEC601, IEC1010, IEC598, IEC990, IEC204...)

  APPLICATION

  MANUFACTURER, LABORATORY, HIGH SPEED PRODUCTION, MAINTENANCE, IN ORDER TO COMPLY WITH THE LOW VOLTAGE DIRECTIVE

  

  • IEC 335 household manufacturer (washing machines, coffee machines, fridges, heaters, boilers, …)

  • IEC 204 industrial equipment (compressor, electrical tools, …)

  • IEC 598 lighting industry (street lights, professional lighting, medical lighting, bulbs, …)

  • IEC 601 medical industry (scanners, surgery equipments, X-ray, electrical beds, …)

  • IEC 950 information technology industry (modems, satellite receivers, servers, printers,cash dispensers, …)

  • IEC 1010 measurement, control and laboratories industry (electrical meters, test & measurement equipment, …)

  SPECIFICATION OVERVIEW

  FROM LABORATORY TO PRODUCTION TESTING, OUR RANGE COVERS THE FOLLOWING SPECIFICATIONS

  • ground bond test up to 64A AC with resistance measurement up to 600mΩ

  • insulation test with DC Voltage from 1V up to 1000 VDC (1V steps),500VA AC Hipot capacity(AC 5KV/100mA),short-circuit output current exceeds 200mA.Built-in optional leakage current test network MD card,meeting requirements of different industrial standards Maximum test current of Ground Bond resistance is up to 64A

  • leakage current measurement in single or three phases, according to all current international standards

  • AC Hipot,Insulation Resistance,Ground Bond resistance,Leakage Current,Running Power,Low-Voltage Start.

  • power measurement and cos φ measurement

  • Satisfy lattest requirements:Follow the production line developing trends of comprehensive,fast,automation and information.Satisfy quality management,5s management and refid management requirements.

  • GPIB, RS232, Ethernet, 0-10V, PLC available

  • ESRS software for quality purposes, traceability and professional reporting



Model
ACW IR GB LC PWR VS Isolated transformer(6kW) AC Power Supply (5kW) Industrial PC ESRS software
AN9651S(F) Yes Yes Yes Yes Yes Yes Yes Built-in Yes Built-in Yes
AN965F(F) Yes Yes Yes Yes Yes Yes Yes  Built-in Optional External Optional
AN9651B(F) Yes Yes Yes Yes Yes Yes Yes
Built-in
OptionalExternal Optional External Optional
AC Hipot test
Booster rated output capacity 500VA(5000V/100mA)
Booster maximum short-circuit current >200mA
Output waveform
distortion and regulation
Sine wave, <3%(pure resistance,5000V/1mA),<3%(From empty load to full load)
Output voltage setting Range 500~5000V
Resolution 5V
Accuracy ±(3%×setting+10V)
Frequency setting Range 50Hz/60Hz
Accuracy ±0.1%×setting
Alarming limits setting Current upper-limit 0.10~99.99mA
Current lower-limit 0.00~10.00mA
Compensation current setting Range 0.00~10.00mA, automatic testing, compensation can be set to ON/OFF
Time setting Range 1~999s
Ramp up time range 1~100s
Accuracy ±(0.1%×setting+1count)
Breakdown current alarm Range 0.10~99.99mA
Accuracy ±(3%×reading+3counts)
Insulation Resistance test
Rated output capacity 2VA(1000V/2mA)
Output voltage regulation <3%(From no load to full load)
Output voltage ripple <1%
Output voltage setting Range DC 500Vor1000V
Accuracy ±(3%×setting+3counts)
Alarming limits setting Resistance upper-limit range 0.0~99.9 MΩ,100~2000MΩ(If upper-limit is set to 0, it will not make adjudgment for upper limit )
Resistance lower-limit range 0.3~99.9 MΩ, 100~2000MΩ
Time setting Range 1~999s
Accuracy ±(0.1%×setting+1count)
Insulation resistance measurement Range 0.0~99.9 MΩ, 100~2000 MΩ
Accuracy ≤200MΩ: ±(3%×reading+3counts)
>200MΩ:±(8%×reading+8counts)
Remarks:Except 500V and 1000V voltage points, other voltage accuracy are all 10%. Insulation resistance accuracy is under temperature 0~40 , humidity 40/(20~75)% RH,otherwise the accuracy will change.
Ground Bond Resistance test
No load output voltage <12V
Output current setting Range 3~30A
Accuracy ±(3%×setting+0.2A)
Current range instruction Optional: 3 ~ 64A test current
Current frequency setting Range Sine waveform 50Hz/60Hz
Accuracy ±0.1%×setting
Test time setting Range 1~999s
Accuracy ±(0.1%×setting+1count)
Result display mode Ground resistance or voltage
Ground resistance display mode:
Alarming limits setting Resistance upper-limit range 3A≤output current≤10A:10~600 mΩ
11A≤output current≤25A:10~300 mΩ
26A≤output current≤30A:10~200 mΩ
Resistance lower-limit range 0~100 mΩ
Compensation resistance setting Range 0~200 mΩ, automatic testing, compensation can be set to ON/OFF
Ground bond resistance measurement Range 10~600 mΩ
Accuracy ±(3%×reading+ 5counts)
Ground voltage display mode:
Alarming limits setting Voltage upper-limit range 30~7500 mV
Voltage lower-limit range 0~3000 mV
Compensation voltage setting Range 0~1000 mV, automatic testing, compensation can be set to ON/OFF
Ground bond resistance measurement Range 10~7500 mV
Accuracy ±(3%×reading+ 50mV)
Leakage Current test
Leakage current test type setting Single phase load, static/dynamic leakage current
MD network selection Standard IEC 60335-1 figure 4 network; other test network optional
Alarming limits Setting Current  upper-limit range 0.050~9.999mA
Current  lower-limit range 0.000~5.000mA
Leakage current compensation setting Range 0.000~1.000mA, automatic testing, compensation can be set to ON/OFF
Time test setting Range 1~999s
Accuracy ±(0.1%×setting+1count)
Leakage current measurement Range 0.050~9.999mA
accuracy DC~10kHz:±(3%×reading+0.010mA)
10k~1MHz:±(5%×reading+0.050mA)
Output voltage measurement Range 60~280V(phase voltage)
accuracy ±(0.5%×reading+2counts)
Run Power  Only for AN9651S(F)/AN9651F(F)/AN9651B(F)
Alarming limits setting Power upper-limit range 6~6000W
Power  lower-limit range 0~6000W
Test time setting Range 1~999s
Accuracy ±(0.1%×setting+1count)
Voltage measurement Range 60~280V(phase voltage)
Accuracy ±(0.5%×reading+2counts)
Current measurement Range 0.030~3.999A, 4.00~25.00A
Accuracy ±(0.5%×reading+2counts)
Active power  measurement Range
Accuracy
30.0~199.9W:±(0.5%×reading+5W)
200~6000W(or2000W):±(0.5%×reading+30W)
Power-factor measurement Range 0.10~1.00
Accuracy ±(2%×reading+2counts)
Voltage Start    Only for AN9651S(F)/AN9651F(F)/AN9651B(F)
Alarming limits setting Current  upper-limit range 0.30~25.00A
Current  lower-limit range 0.00~25.00A
Test time setting Range 1~999s
Accuracy ±(0.1%×setting+0.05s)
Voltage  measurement Range 60~280V(phase voltage)
Accuracy ±(0.5%×reading+2counts)
Current measurement Range 0.030~3.99A,4.00~25.00A
Accuracy ±(0.5%×reading+2counts)
DUT power control
Isolation transformer input voltage, frequency 0~300V(phase voltage), 45~65Hz
Isolation transformer output voltage ratio Leakage current test 1.06 times  rated voltage
Running Power test 1.00 times rated voltage (AN9641B(F) without output)
Voltage start test 0.85 times  rated voltage (AN9641B(F) without output)
Isolation Transformer Capacity Single phase, rated capacity is 6 KVA
Isolation transformer voltage regulation 3%(From empty load to full load)
Over current protection Max current 30A, after the over current 5S,automatically cut off the isolation transformer
AN9651B(F) can be combine with AN97 series AC frequency conversion power supply, more accurate. Realize voltage stabilization, voltage regulation and frequency conversion. AN9651S(F)/AN9651F(F) have been fixed 5KW AC frequency conversion power supply in the case.
 
Interface
Remote interface Standard, foot pedal(start/stop)
Alarming lamp interface Standard, 3 color light(test\pass\fail)
Communication interface Standard RS232,internet, RS485, GPIB optional
97 Power supply control interface Optional,combine with AN97 series frequency conversion power
PLC interface Optional, start、stop、test、pass、fail output
Printer interface Optional
General specification
Operating environment 0~40℃, 40℃/(20~75)% RH, little dust
Storage environment -40~60℃ , 50℃/(20~90)% RH, little dust
Input power(DUT consumption is not included ) 220V±10%, 50Hz±5%, 10A
Power consumption(DUT consumption is not included ) 100~600VA(According to working condition)
Gross Weight About 150kg
Dimension With packing AN9651S(F) 605(W)x1640(H)x780(D)
AN9651F(F) 605(W)x1310(H)x780(D)
AN9651B(F) 605(W)x1270(H)x666(D)
Without  packing AN9651S(F) 483(W)x 1355(H)x 600(D)
AN9651F(F) 483(W)x 1105(H)x 600(D)
AN9651B(F)/ 530(W)x 1105(H)x 460(D)


 India  Malaysia  Indonesia
 Brazil  Venezuela  Thailand
 Poland  Hungary  Czech Republic
 Bangladesh  Sri Lanka  Vietnam
 Turkey  Egypt  Pakistan
 Korea  Australia  New-Zealand
 Russia  Saudi Arabia  Dubai
 United States  Canada  Mexico
 Nicaragua  Haiti Columbia
 Ecuador  Peru  Uruguay
Argentina  Chile  Italy
Austria  Czech Rep  Belgium
Netherlands United Kingdom Ireland
France Spain Germany
Portugal Finland  

Product:
Country:
Company:
Name:
Phone:
Email:
Message:
 
Products / Products